Acceptance of ET_8 tagger in 1998-1999.
V.Andreev, 26-April-2000
The new electron tagger ET8 was installed in the HERA tunnel
in 1998 at the distance 8 m from the interaction point. This
detector effectively accepts the scattered electrons with energy
several GeV that gives opportunity to tag the photoproduction
processes with the higher possible photon energy Wgp=300 GeV.
More details about the calorimeter design, performance,
procedure of acceptance determination and obtained results during
1998-1999 HERA Luminosity Runs are described in the H1 note.
The acceptance of the ET8 is determinated with the same
procedure which is used at the analyses of ET33 and ET44 taggers.
The information about the acceptance behaviour and the trigger
efficiency for the different sub-samples is stored in the new (run-
dependent) bank LET2 in the H1 database. The structure of this
bank is shown below
! BANKname BANKtype !Comments
!
TABLE LET2 ! ET_8 acceptance parametrization
!
! ATTributes:
! -----------
!COL ATT-name FMT Min Max !Comments
!
1 EFF0 F 0. 1. ! Trigger efficiency
2 YM F 0. 1. ! Trigger threshold value
3 SIGY F 0. 1. ! Sigma of trigger threshold
4 PVETO F 0. 1. ! Gamma-arm veto level
5 YMIN F 0. 1. ! Minimum y-value
6 YMAX F 0. 1. ! Maximum y-value
7 DY F 0. 1. ! =0.005 dy-step for parametrization
8 Y1 F 0. 1. !
... !
N+8 YN F 0. 1. !
END TABLE
The simulation of the ET8 is not presently included in H1SIM
and for this reason it is recomended to use the new ACET8(IRUN,Y,MODE)
routine both for the analysis of the experimental data and for the
Monte Carlo events. This routine provides the acceptance value for
the different run periods and for the given y-value. Chose a proper
option MODE the acceptance dependence on the trigger efficiency
and on the gamma-arm veto level could be taken into account.
The systematic error is formed by the next sources:
a) precision of the electron energy scale --> 4.5 % ,
b) uncertainty in the ET-8 position with respect
to the electron beam --> 4.0 % ,
c) trigger efficiency determination --> 1.0 % ,
d) fitting procedure and averaging of the
calibration samples --> 1.0 % .
So the overall systematic error in the acceptance determination
of the ET-8 in 1999 could be taken as 6.2 % .